Criteria for Selection of a Bandelin Sonorex Ultrasonic Bath.

  SONOREX DIGITEC SONOREX SUPER
Tank volume (litres) 0,9 – 90,0 0,9 – 58,0
Control elements push-buttons turning knobs
Time setting (min) 1– 30, continuous operation ∞ 1–15, ontinuous operation ∞
Safety shut-down after 12 hours no
Heater optional, version „H“ optional, version „H“
Heater, thermostaically adjustable 20 – 80°C 30 – 80°C, RK 31 H: 65 °C fixed
Excess temperature signal yes no
Protection against delay in boiling yes, optionally switch-on no
Setting accuracy of bath temperature ± 3,5 K ± 5 K
Thickness of s/s tank/material 0,8 mm, 1.4301 0,8 mm, 1.4301
Marking of fi lling level for safe dosage yes yes
Onoe-piece drain yes, from DT 255 yes, ab RK 255
Liquid protection protected against spray drip-proof
Protection class IP 33 IP 32
Ultrasonic frequency (kHz) 35 35
Sweep yes yes
PZT-transducers yes yes
Degas yes no
Mains supply 230 V~, 50/60 Hz yes yes
Data memory 1 program no
CE marked as medical device yes yes